EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4445A

Agilent Technologies E4445A Manual

Agilent Technologies E4445A
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #249 background imageLoading...
Page #249 background image
Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
249
Front-Panel Key and SCPI Command
Reference
4.4.9.1.3 Ratio Reference
If AM depth is to be displayed relative to a reference, enter the value as a ratio reference using this key.
4.4.9.1.4 Ratio Mode
The ratio and previous ratio functions permit any measurement result to be scaled to a reference. The reference
may be either a measurement result or a keyboard entry.
The Log/Linear key enables measurement results to be displayed in linear or logarithmic units. For purposes of
discussion, % for AM is treated as absolute unit. When the instrument display dB as relative unit, annunciator
REL is added to the display.
Key Path: Amplitude Y Scale
Remote Command: :DISPlay:MODSinad:RREFerence <rel_ampl>
:DISPlay:MODSinad:RREFerence?
:DISPlay:MODSinad:RREFerence:AUTO 0|1|OFF|ON
:DISPlay:MODSinad:RREFerence:AUTO?
Preset: 0.00 dB
State Saved: Saved in instrument state.
Min: 0.00 dB
Max: 200.00 dB
Example: :DISP:MODS:RREF 10 dB
:DISP:MODS:RREF?
:DISP:MODS:RREF:AUTO OFF
:DISP:MODS:RREF:AUTO?
Key Path:
Amplitude Y Scale
Remote Command: :DISPlay:MODSinad:RMODe LOG | LINear
:DISPlay:MODSinad:RMODe?
Preset: LOG
State Saved: Saved in instrument state.
Range: Log | Linear
Example: :DISP:MODS:RMOD LOG
:DISP:MODS:RMOD?

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals