Chapter 3 109
Making Measurements
Modulation SINAD Measurement
Making Measurements
3. Press FM De-Emphasis to select a value. This key is enabled only when the
Modulation Mode is FM.
4. Press
Modulation Mode to select from AM, FM and PM.
Step 9. To adjust the display, press
AMPLITUE/Y Scale, then:
1. Press the
Display Unit key and select a display unit key. The default setting
is
dB.
2. Press the
Display Mode key to toggle between Normal and Ratio. The default
setting is
Normal.
3. Press the
Ratio Ref key to toggle between Auto and Man. When you select Man,
input the reference value and unit. When the reference is set to 0 or the
measurement result is 0 under log ratio mode, the result displayed is ā€œ- - -ā€.
4. Press the
Ratio Mode key to toggle between Log and Linear.
Modulation SINAD Measurement Example
Figure 3-20 AM Modulation SINAD Measurement Result
Signal Under Test: RF Frequency: 1.25 GHz; Amplitude: –30.0 dBm;
AM Depth: 10%; AM Rate: 100 kHz; Modulation: On.
In this example, the measured AM Modulation SINAD is 39.665 dB.

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals