Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
137
Front-Panel Key and SCPI Command
Reference
4.2.3 Power Sensor
Accesses the menu for power sensor configurations. This key is grayed out when current measurement is one of
the four Audio measurements: Audio Frequency, Audio AC Level, Audio Distortion and Audio SINAD.
When the PSA is powered-on, the retained data will be displayed.
4.2.3.1 Add New Points
Adds new points.
4.2.3.2 Calibration Due Date
Sets the Calibration Due date of the sensor
4.2.3.3 Calibration Due Date String
Gets the Calibration Due date of the sensor in the type of string
Mode: MRECEIVE
Remote Command: :SYSTem:CFACtor:NPS <real>,<real>{,<real>,<real>}
Preset: 1e9, 100
State Saved: No
Remote Command Notes: Command Only
Example: :SYSTem:CFAC:NPS 1.5e9,98.876,1.34e6,92.34562,100,87.3
Mode: MRECEIVE
Key Path:
System, Power Sensor Config
Remote Command: :SYSTem:PSENsor:DDATe <Integer>
:SYSTem:PSENsor:DDATe?
Preset: 20000101
State Saved: Saved in instrument state.
Min: 19000101
Max: 99991231
Example: :SYSTem:PSEN:DDAT 20120803
:SYSTem:PSEN:DDAT?
Mode: MRECEIVE
Remote Command: :SYSTem:PSENsor:DDATe:STRing?
Preset: 20050101 00:00:00 AM
State Saved: Saved in instrument state.
Example: :SYSTem:PSEN:DDAT:STR?

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals