Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
126
Front-Panel Key and SCPI Command
Reference
4.2.1.1.1 Peak Hold Turns on/off the Peak Hold mode.
4.2.1.2 High Pass Filter
Sets the type of high pass filters used by the instrument. The frequency indicated is the low frequency cut-off.
Mode: MRECEIVE
Key Path:
Det/Demod, Detector
Remote Command: [:SENSe]:MRECeive:PHOLd ON|OFF|0|1
[:SENSe]:MRECeive:PHOLd?
Preset: OFF
State Saved: Saved in instrument state.
Range: On | Off
Dependencies/Couplings: This key is grayed out when the selected Detector is not either Peak + or Peak
−. In this case, the Peak Hold switch is forced to be OFF.
Remote Command Notes: Global to the current mode.
Example: MREC:PHOL ON
MREC:PHOL?
Mode: MRECEIVE
Key Path:
Det/Demod
Remote Command: [:SENSe]:MRECeive:HPFilter OFF|F50|F300
[:SENSe]:MRECeive:HPFilter?
Preset: OFF
State Saved: Saved in instrument state.
Range: None (<20 Hz) | 50 Hz | 300 Hz
Dependencies/Couplings: For Audio measurements the None (<20 Hz) key will be displayed as None,
because the spec of Audio HW is from 20 Hz to 250 kHz.
Remote Command Notes: Global to the current mode.
Example: MREC:HPF F50
MREC:HPF?

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals