Chapter 4
Front-Panel Key and SCPI Command Reference
Front-Panel Keys
128
Front-Panel Key and SCPI Command
Reference
4.2.1.5 FM De-Emphasis
Sets the types of de-emphasis filters.
4.2.1.6 Modulation Mode
Sets the types of analog modulation.
Mode: MRECEIVE
Key Path:
Det/Demod
Remote Command: [:SENSe]:MRECeive:DEEMphasis OFF|T25|T50|T75|T750
[:SENSe]:MRECeive:DEEMphasis?
Preset: OFF
State Saved: Saved in instrument state.
Range: None | 25 us | 50 us | 75 us | 750 us
Dependencies/Couplings: This key is only available when Modulation Mode is set to FM or the current
measurement is one of the Audio measurements.
Remote Command Notes: Global to the current mode.
Example: MREC:DEEM T25
MREC:DEEM?
Mode: MRECEIVE
Key Path:
Det/Demod
Remote Command: [:SENSe]:MRECeive:MODulation:TYPE AM|FM|PM
[:SENSe]:MRECeive:MODulation:TYPE?
Preset: AM
State Saved: Saved in instrument state.
Range: AM | FM | PM
Dependencies/Couplings: This key is grayed out when current measurement is one of the four Audio
measurements.
If the current measurement is AM/FM/PM, the value must be AM/FM/PM
and can not be modified.
Remote Command Notes: Global to the current mode.
Example: MREC:MOD:TYPE FM
MREC:MOD:TYPE?

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals