Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
233
Front-Panel Key and SCPI Command
Reference
4.4.6.4.6 Capture Time
Sets the capturing data time. When in manual mode, the user determines the capture time; otherwise, it is
determined automatically by the measurement.
This key is available only when PSA firmware revision A.11.08 or above, or Option 23A is installed.
4.4.6.4.7 Fast Mode
Allows you to toggle the
Fast Mode between On and Off. Fast Mode can speed up the measurement.
Key Path: Meas Setup
Remote Command: [:SENSe]:PMDeviation:CAPTure[:TIME] <time>
[:SENSe]:PMDeviation:CAPTure?
[:SENSe]:PMDeviation:CAPTure:AUTO 0|1|ON|OFF
[:SENSe]:PMDeviation:CAPTure:AUTO?
Preset/Default: 250 ms
State Saved: Saved in instrument state.
Min: 1 μs
Max: 10 s
Resolution 1 ms
Example: :PMD:CAPT 0.25
:PMD:CAPT?
:PMD:CAPT:AUTO OFF
:PMD:CAPT:AUTO?
Key Path:
Meas Setup
Remote Command: [:SENSe]:PMDeviation:FAST 0|1|OFF|ON
[:SENSe]:PMDeviation:FAST?
Default: OFF
State Saved: Saved in instrument state.
Range: ON|OFF
Example: :PMD:FAST ON
:PMD:FAST?