362 Chapter 7
Using GPIB Power Meter in the System
Using a GPIB Power Meter in the System
Using GPIB Power Meter in the System
Step 6. Press the Power Meter GPIB Address key and enter the GPIB address of your
power meter. To determine the power meter GPIB address on the Agilent E4418 or
E4419A power meters, press
System/Inputs, GPIB on the power meter front panel.
Step 7. Press the
Verify Power Meter Connection key. The grayed-out Show Setup key will
be enabled when the connection between the PSA and the power meter is
established.
Step 8. Select the power meter channel number you would like to use by toggling the
Channel key. If the power meter has only one channel, this key is grayed out.
Step 9. Press the
Show Setup key to display and verify the power meter configuration
information.
Step 10. You can set the desired time-out for your testing by pressing
Power Meter Config,
More, Time Out keys.
Step 11. Press
Power Meter Config, More, Resolution to select the resolution as 0.001 dB
or 0.01 dB.

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals