8-14 Service Guide E8364-90026
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers
Error Terms E8362B, E8363B, E8364B
Isolation (E
XF
and E
XR
)
Isolation, or crosstalk, is the uncorrected forward and reverse isolation error terms that
represent leakage between the test ports and the signal paths. The isolation error terms
are characterized by measuring transmission (S
21
, S
12
) with loads attached to both ports
during the measurement calibration.isolation errors affect transmission measurements
primarily where the measured signal level is very low.
The measurements most affected by isolation error terms are DUTs with large insertion
loss. Since these terms are low in magnitude, they are usually noisy (not very repeatable).
Typical Cause of Failure A loose cable connection or leakage between components in the
test set are the most likely cause of isolation problems.
After verifying the cable and its connections, suspect the analyzer switch splitter, step
attenuator, coupler, or receivers, and associated cabling.
To troubleshoot, refer to “Checking the Receiver Group” on page 4-52 or to “Checking the
Signal Separation Group” on page 4-48.

Table of Contents

Other manuals for Agilent Technologies E8362C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8362C and is the answer not in the manual?

Agilent Technologies E8362C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8362C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals