EasyManuals Logo

Agilent Technologies E8362C User Manual

Agilent Technologies E8362C
392 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
3-46 Service Guide E8364-90026
Tests and Adjustments PNA Series Microwave Network Analyzers
Adjustments E8362B, E8363B, E8364B
6. Connect the power sensor 8487A and the 11901D adapter or the E4413A power sensor
to the end of the right angle adapter as shown in Figure 3-19 on page 3-43.
7. Reconnect the power and USB cables. Turn on the analyzer.
8. On the System menu, point to Service, Adjustments, and then click LO Power Adjust.
9. Ensure the GPIB settings are correct.
10. Click Begin Adj, and then follow the instructions as they are displayed.
11. Turn off the analyzer, disconnect the power cable, reconnect the semi-rigid cables, and
replace the covers.

Table of Contents

Other manuals for Agilent Technologies E8362C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8362C and is the answer not in the manual?

Agilent Technologies E8362C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8362C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals