7-26 Service Guide E8364-90038
Repair and Replacement Procedures PNA Series Microwave Network Analyzers
Removing and Replacing the A14 System Motherboard E8362C, E8363C, E8364C
Removing and Replacing the A14 System Motherboard
Tools Required
T-10 TORX driver (set to 9 in-lb)
T-20 TORX driver (set to 21 in-lb)
5/16-inch open-end torque wrench (set to 10 in-lb)
ESD grounding wrist strap
Removal Procedure
Refer to Figure 7-11 for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to “Removing the Covers” on page 7-6.
3. Remove the front panel assembly. Refer to “Removing and Replacing the Front Panel Assembly” on
page 7-10.
4. Remove the A12 source 20. Refer to “Removing and Replacing the A12 Source 20” on page 7-22.
5. Remove the A13 (Option 080) frequency offset receiver assembly, if present. Refer to “Removing and
Replacing the A13 Frequency Offset Receiver Assembly” on page 7-24.
6. Remove the A11 phase lock board. Refer to Removing and Replacing the A11 Phase Lock Board” on
page 7-20.
7. Remove the A4 power supply assembly. Refer to “Removing and Replacing the A4 Power Supply
Assembly” on page 7-16.
8. Remove the card cage boards (A6, A8, and A10). Remove A9 (Option 080), if present. Refer to “Removing
and Replacing the A6, A8, A9, and A10 Boards” on page 7-18.
9. Remove the A15 CPU board. Refer to “Removing and Replacing the A15 CPU Board” on page 7-28.
10. Remove the Midweb. Refer to “Removing and Replacing the Midweb and the B1 Fan” on page 7-76.
11. Remove the ribbon cable clamp (item
) by removing the attachment screw (item
). Set the cable
clamp aside for reinstallation later.
12. Disconnect the front panel ribbon cable (item
), the test set motherboard ribbon cable (item ), and
the USB hub assembly cable (item
) from the A14 system motherboard.
13. Remove the remaining three screws (item
) that secure the A14 system motherboard.
14. Lift the A14 system motherboard out of the analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain to this removal
procedure. Refer to Table 7-2 on page 7-82.

Table of Contents

Other manuals for Agilent Technologies E8363C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8363C and is the answer not in the manual?

Agilent Technologies E8363C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8363C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals