Service Guide E8364-90038 8-21
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures
E8362C, E8363C, E8364C Correction Constants
Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM memory. These constants enable
the analyzer to produce accurate, leveled source signals and receive clean test signals.
Storing Correction Constants
After performing any adjustment listed on page 3-38 in this manual, store the correction constants to a
backup file on the analyzer hard disk drive by performing these steps:
• Navigate to the EEPROM Backup Utility, located at:
C:\Program Files\Agilent\Network Analyzer\Service\eebackup.exe
•
Run the program.
•Click
Backup EEPROM.
•Click
Exit when the program has finished.