Service Guide E8364-90038 3-23
PNA Series Microwave Network Analyzers Tests and Adjustments
E8362C, E8363C, E836C System Verification
Precautions for Handling Airlines
The verification kit includes airlines. Follow the steps below to ensure good measurement techniques.
• Be very careful not to drop the airline’s center or outer conductor. Irreparable damage will result if these
devices are dropped.
• During this procedure, you will be touching the exposed center conductor of the test port and the center
conductor of the airline.
— Observe proper ESD procedures.
— Clean your hands or wear gloves; skin oils will cause a change in electrical performance.
• Refer to Figure 3-9, Figure 3-10, and to your calibration kit manual for connection procedures.
Figure 3-9 Aligning the Center Conductor
Figure 3-10 Torquing the Connection