Installing the G1330B Thermostat
Transporting the Thermostatted Autosampler
Transporting the Thermostatted
Autosampler
When moving the thermostatted autosampler around the laboratory, make
sure that any condensed water inside the thermostat is removed. Tilt the
module to the front, so that the water inside the thermostat can safely flow
into the leak funnel. Otherwise no special precautions are needed for the
modules.
NOTE The ALS thermostat is heavy (20.7 Kg, 45.6 lbs). Carry the module, by putting
your hands under the side covers in a central position of the unit.
If the thermostatted autosampler needs to be shipped to another location via
carrier, ensure:
The two modules are shipped in separate boxes.
The transport assembly of the autosampler is parked, see Park Arm (Park
Gripper) in your respective Reference Manual.
The vial tray is secured.
If the thermostatted autosampler is to be shipped to another location, the
transport assembly of the autosampler must be moved to the park position to
prevent mechanical damage should the shipping container be subjected to
excessive shock. Also, ensure the vial tray is secured in place with suitable
packaging, otherwise the tray may become loose and damage internal
components.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1330B and is the answer not in the manual?

Agilent Technologies G1330B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1330B
CategoryThermostat
LanguageEnglish

Related product manuals