Repairing the Autosampler
Needle-Seat Assembly
6 Ensure the needle is aligned with the seat. If
required, bend the needle slightly until the
needle is aligned correctly.
On completion of this procedure:
Install the front cover.
Select End in the maintenance function
Change Needle (see Change Needle on
page 81).