EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>G2260A

Agilent Technologies G2260A Reference Manual

Agilent Technologies G2260A
292 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
77
Troubleshooting and Test Functions
Vial in Gripper
Vial in Gripper
The gripper arm attempted to move with a vial still in the gripper.
During specific stages of the sampling sequence, no vial should be held by the
gripper. The autosampler checks if a sample vial is stuck in the gripper by
closing and opening the gripper fingers while monitoring the motor encoder.
If the gripper fingers are unable to close, the error message is generated.
Probable Causes Vial still in gripper.
Suggested Actions Remove the vial using the Release Vial function in the user interface.
Reinitialize the autosampler.

Table of Contents

Other manuals for Agilent Technologies G2260A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G2260A and is the answer not in the manual?

Agilent Technologies G2260A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG2260A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals