Troubleshooting and Test Functions
Arm Movement Failed
Arm Movement Failed
The transport assembly was unable to complete a movement in one of the
axes.
The processor defines a certain time window for the successful completion
of a movement in any particular axis. The movement and position of the
transport assembly is monitored by the encoders on the stepper motors. If
the processor does not receive the correct position information from the
encoders within the time window, the error message is generated.
See figure Transport Assembly on page 210 for axes identification.
Arm Movement 0 Failed: X-axis.
Arm Movement 1 Failed: Z-axis.
Arm Movement 2 Failed: Theta (gripper rotation).
Arm Movement 3 Failed: Gripper (gripper fingers open/close).
Probable Causes Mechanical obstruction.
High friction in transport assembly.
Defective motor assembly.
Defective transport assembly flex board.
Defective ASM board.
Suggested Actions Ensure unobstructed movement of the transport assembly.
Exchange the transport assembly.
Exchange the ASM board.

Table of Contents

Other manuals for Agilent Technologies G2260A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G2260A and is the answer not in the manual?

Agilent Technologies G2260A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG2260A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals