82
Troubleshooting and Test Functions
Change Needle
3 Remove the front cover.
NOTE Do not remove the front cover until the needle arm is in its maintenance
position. Removing the cover while the needle arm is activated may lock up
the system.
4 Exchange the needle or needle seat (see Needle-Seat Assembly on page 111
and Needle Assembly on page 108).
5 Replace the front cover.
6 Select End (Done) to complete the procedure.