20 Agilent ADM Flow Meter Operating Instructions
For information on contacting Agilent Technical Support, see “Technical
Support" on page 30.
0007 Warning Measurement exceeds
instrument range.
Measure gas flows within
the acceptable flow range.
0008 Warning Cartridge calibration
expired.
Install a new cartridge.
0009 Warning RTC (real time clock)
backup battery low.
Contact Agilent Technical
Support.
0010 Warning Out of operating tempera-
ture range condition (0 -
45
C
Operate the flow meter
within the acceptable tem-
perature range.
0011 Warning Out of operating atmo-
spheric pressure range.
Operate the flow meter
within the acceptable atmo-
spheric pressure range.