6.4.1 Check water and argon gas
(1) If error 1440 has been generated, check if cooling water valve is open and chiller or heat
exchanger is working correctly.
(2) If error 1445 or 1446 has been generated, check if argon gas valve is open and remaining
amount is enough.
(3) Did this solve the problem
Yes No