Agilent InfiniiVision 3000 X-Series Oscilloscopes
User's Guide
s1
12
Acquisition Control
Running, Stopping, and Making Single Acquisitions (Run Control) 185
Overview of Sampling 187
Selecting the Acquisition Mode 191
Acquiring to Segmented Memory 197
This chapter shows how to use the oscilloscope's acquisition and run
controls.
Running, Stopping, and Making Single Acquisitions (Run Control)
There are two front panel keys for starting and stopping the oscilloscope's
acquisition system: [Run/Stop] and [Single].
When the [Run/Stop] key is green, the oscilloscope is running, that is,
acquiring data when trigger conditions are met.
To stop acquiring data, press [Run/Stop]. When stopped, the last
acquired waveform is displayed.
When the [Run/Stop] key is red, data acquisition is stopped.
"Stop" is displayed next to the trigger type in the status line at the top
of the display.
To start acquiring data, press [Run/Stop].
To capture and display a single acquisition (whether the oscilloscope is
running or stopped), press [Single].
The [Single] run control lets you view single- shot events without
subsequent waveform data overwriting the display. Use [Single] when
you want maximum memory depth for pan and zoom.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 DSO-X 3024A and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 DSO-X 3024A Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 DSO-X 3024A
CategoryTest Equipment
LanguageEnglish

Related product manuals