410 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
27 UART/RS232 Triggering and Serial Decode
Rx Stop Bit — Triggers when a stop bit occurs on Rx. The trigger will
occur on the first stop bit. This is done automatically whether the
device under test uses 1, 1.5, or 2 stop bits. You do not need to
specify the number of stop bits used by the device Under test.
Rx Data — Triggers on a data byte that you specify. For use when the
device under test data words are from 5 to 8 bits in length (no 9th
(alert) bit).
Rx 1:Data — For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 1. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
Rx 0:Data — For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers only when the
9th (alert) bit is 0. The specified data byte applies to the least
significant 8 bits (excludes the 9th (alert) bit).
Rx X:Data — For use when the device under test data words are 9 bits
in length including the alert bit (the 9th bit). Triggers on a data byte
that you specify regardless of the value of the 9th (alert) bit. The
specified data byte applies to the least significant 8 bits (excludes
the 9th (alert) bit).
Similar choices are available for Tx.
Rx or Tx Parity Error — Triggers on a parity error based on the parity
you have set in the Bus Configuration Menu.
6 If you choose a trigger condition that includes "Data" in its description
(for example: Rx Data), then press the Data is softkey, and choose an
equality qualifier. You can choose equal to, not equal to, less than, or
greater than a specific data value.
7 Use the Data softkey to choose the data value for your trigger
comparison. This works in conjunction with the Data is softkey.
8 Optional: The Burst softkey lets you trigger on the Nth frame (1- 4096)
after an idle time you select. All trigger conditions must be met for the
trigger to occur.
9 If Burst is selected, an idle time (1 µs to 10 s) can be specified so that
the oscilloscope will look for a trigger condition only after the idle time
has past. Press the Idle softkey and rotate the Entry knob to set an idle
time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 DSO-X 3024A and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 DSO-X 3024A Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 DSO-X 3024A
CategoryTest Equipment
LanguageEnglish

Related product manuals