Key Reference 5
N9342C/43C/44C User’s Guide 129
• Zero Type
Toggles the zeroing type of the Power Sensor
between internal and external. When this type is
set to EXT, pressing Zeroing will bring up a
message which indicates the external signal
need be pulled out. This setting is not affected
by the preset in Power Meter menu.
Meas Disp
Key access: [MODE] > {Power Meter}
Access the Meas Disp submenu for defining the
display range and resolution while in Power Meter
mode.
• Disp Range
Access the associated Disp Range submenu.
• Auto Range
Toggles the auto range function between On and
Off. When the Auto Range is set to On, the Top
and Bottom value of the power meter’s display
range will be set automatically according to the
measurement data:
Top = Max. data + 0.5
Bottom = Min. data - 0.5
The default setting is Off.
• To p
Set the top value of the power meter’ display
range.
Range from (Bottom + 0.01) to 200 dBm.
The default value is 20 dBm.
• Bottom
Set the bottom value of the power meter’ display
range.
Range from - 200 to (Top - 0.01) dBm.
The default value is - 70 dBm.