2 Getting Started
22 N9342C/43C/44C User’s Guide
Timed Power On/Off
Pressing [System] > {Power On/Off Preset} > {Timed Pwr
On} or {Timed Pwr Off} sets the time switch to power
on/off the N934xC in a user- defined time and date.
This function requires the power supply to be
connected or charged battery installed.
Press {Repeat Mode Once/Everyday} to set the N934xC
boot up/off in the pre- saved time everyday. The
pre- saved date is invalid in this mode.
IP configuration
The N934xC supports LAN port connection for
data transfer. Press [System] > {Port Setting} > {IP
Admin} > {IP Address Static} to manually set the IP
address, gateway and subnet mask with the proper
LAN information. Or, just press [System] > {Setting}
> {IP Admin} > {IP Address DCHP} to get the IP address
in LAN dynamically according DCHP.
Press {Apply} to enable all the configurations you
set.
Ext Input
Toggles the channel for external input between Ref
and Tr ig . Ref refers to a 10 MHz reference signal; Tri g
refers to a TTL signal.
External Reference (Ref)
Use the external reference function as follows:
1 Input a 10 MHz signal into the EXT TRIG IN/REF IN
connector.
2 Press [System] > {Port Setting} > {Ext Input Ref} to
enable the external reference signal input.
Then, the analyzer will disable the internal
reference and switch to accept the external
reference.