3 Functions and Measurements
78 N9342C/43C/44C User’s Guide
frequency domain reflectometry. The transmitted
and reflected signals contain information about the
distance to the faults. This information is used to
determine the physical distance to the faults. The
tester displays the physical distance to the
probable faults, corrected for cable loss and the
velocity propagation factor of the cable.
Perform a DTF measurement
To make a DTF- return loss measurement over a
frequency range of 50 to 400 MHz, perform the
following steps:
1 Press the [MODE] > {Cable&Aetenna Test}>{Distantce
To F a ul t} to select the DTF mode.
2 Press [FREQ] > {Start} > 50 > [ENTER] to set the
start frequency to 50 MHz.
3 Press {Stop} > 400 > [ENTER] to set the stop
frequency to 400 MHz.
4 Press {Calibrate} to bring up the calibration
process guidance. Follow the instructions to
perform a calibration. For more information
about calibration, refer to “Preparation" on
page 76.
5 Remove the calibrator and connect the cable
under test (DUT) to the RF OUT connector on
the top panel of the tester.
The DTF measurement starts and the results are
displayed on the screen. A peak in the horizontal
trace indicates a defective location.

Table of Contents

Other manuals for Agilent Technologies N9344C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9344C and is the answer not in the manual?

Agilent Technologies N9344C Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9344C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals