Technical Information
J1 – IN-OUT
166/192 TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01
For example:
reference quantity: frequency
threshold: 867 Hz
hysteresis: 2 %
activation type: "high level"
Figure 20
It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
reference quantity: frequency
threshold: 867 Hz
hysteresis: 2 %
activation type: high level
delay time: 0 second
NOTE
The A-PLUS (optional) allows the operator to set all the programmable features.