Technical Information
Description
Window Meanings
N. Read/ Write Data
Type
Description Permitted Values
000 R/W L Start/Stop
(in remote/ mode the window is read only)
Start = 1
Stop = 0
008 R/W L Remote
or Serial configuration
Remote = 1
Serial = 0
(default = 1)
100 R/W L Soft Start
(write only in Stop condition)
YES = 1
NO = 0
(default = 0)
101 R/W N R1 Set Point type 0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 0)
102 R/W N R1 Set Point value (expressed in Hz, W or s) (default = 867)
103 R/W N Set Point delay: time between pump start and
the set point check (seconds)
0 to 999999
(default = 0)
104 R/W L Set Point signal activation type: the signal can
be "high level active" or "low level active"
0 = high level active
1 = low level active
(default = 0)
105 R/W N Set point
Hysteresis
(in % of value)
0 to 100
(default = 2)
TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01 181/192