Technical Information
Description
17
Window Meanings
N. Read/ Write Data
Type
Description Permitted Values
000 R/W L Start/Stop
(in remote/ mode the window is read only)
Start = 1
Stop = 0
008 R/W L Remote
or Serial configuration
Remote = 1
Serial = 0
(default = 1)
100 R/W L Soft Start
(write only in Stop condition)
YES = 1
NO = 0
(default = 0)
101 R/W N R1 Set Point type 0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 0)
102 R/W N R1 Set Point value (expressed in Hz, W or s) (default = 867)
103 R/W N Set Point delay: time between pump start and
the set point check (seconds)
0 to 999999
(default = 0)
104 R/W L Set Point signal activation type: the signal can
be "high level active" or "low level active"
0 = high level active
1 = low level active
(default = 0)
105 R/W N Set point
Hysteresis
(in % of value)
0 to 100
(default = 2)
TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01 181/192

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies TwisTorr 74 FS AG and is the answer not in the manual?

Agilent Technologies TwisTorr 74 FS AG Specifications

General IconGeneral
BrandAgilent Technologies
ModelTwisTorr 74 FS AG
CategoryController
LanguageEnglish

Related product manuals