Technical Information
Description
N. Read/ Write Data
Type
Description Permitted Values
106 R/W L Water cooling (write only in stop) 0 = NO
1 = YES
107 R/W L Active Stop (write only in stop) 0 = NO
1 = YES
(default = 0)
108 R/W N Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
109 W L Pump life/ cycle time/ cycle number reset
To reset, write ‘1’
110 R/W L Interlock type
(default = 1)
Impulse = 0
Continuous = 1
111 R/W L Analog output type: output voltage signal
proportional to frequency or power
0 = frequency
1 = power
(default = 0)
2 = Pump temperature
3 = Pressure
120 R/W N Rotational frequency setting (Hz) 1100 to 1167 (default
= 1167)
122 R/W L Set vent valve on/off
0 = Closed
1 = Open
(default = 1)
125
R/W
L
Set the vent valve
operation
Automatic = 0
(see note 1.)
On command = 1
(see note 2.)
(default=0)
126 R/W N Vent valve opening
delay
(expressed in 0.2 sec)
0 to 65535
(corresponding to 0
to 13107 sec)
(default = 15)
182/192 TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01