28 03-914852-00:4
Automation Page Parameters
The CP-8400/8410 has a Single Method automation mode and a Method
Sequence mode. The Single Method Automation consists of one page if
the Setup parameter Number of Injection Ports is set to either Position 1
or Position 2, or two pages if the Number of Injection Ports in Setup is set
to Both. Similarly, the Method Sequence automation mode requires one
page for a single defined injector and two pages for both injectors. A
Priority Sample ‘override’ is provided for urgent analyses.
Single Method Automation Parameters
Page1
Method
Method 1 - Method 8 default = Method 1
Initial Sample
0 to 99 (1 to 21)* default = 0
Final Sample
0 t o 99 (1 to 21)* default = 99
Injections / Sample
1 to 99 default = 99
First Injection Volume (µ
µµ
µL)
0.1-5, 0.1-10, or 1-100 default = 1.0
Second Injection Volume (µ
µµ
µL)
0.1-5, 0.1-10, or 1-100 default = 1.0
(if a single injector is configured, only the first Injection Volume will be displayed)
(if SPME is being used, the Injection Volume fields will have dashed lines)
Page 2 (if Both selected in Setup)
Use Prep-Ahead?
Yes or
No default = No
Injector Position
Pos. 1, Pos. 2, Both default = Pos. 1
First Injector Used
Pos. 1, Pos. 2 default = Pos. 1
Use Injection Delay?
Yes or No default = No
Delay Between Injections (min)
0.5-10.0 default = 1
Advance for Second Injection?
Yes or No default = Yes
Clean Between Injections?
Yes or No default = No
* values in parentheses denote CP-8410 limits
Note that the Page 2 Parameters are only used if Both is selected
for the Inj Pos parameter.
Under Workstation control, user access and control of automation
through the Local User Interface is not allowed; viewing the existing
Automation setup parameters is allowed. Workstation control will function
through its own SampleList and automation structures.