8 03-914852-00:4
Syringe Installation
10. If the currently installed CP-3800 ROM is version 3.1 or lower,
replace with a version 3.3.5 or higher ROM board. To use the
CP-8400 with the CP-3800 Gas Chromatograph, your CP-3800
must have Version 3.2 or later software installed. The CP-8410
requires version 3.3 or later. To use SPME mode sample
introduction with either the CP-8400 or CP-8410 you must have
Version 3.3.5 or later software installed.
11. Replace the left side top cover.
12. Place a vial with cap into the Vial 0 location if you have a
CP-8400 or the Vial 1 location if you have a CP-8410.
13. Place the CP-8400 carrousel onto the base or place the
CP-8410 sample tray oriented so that Vial 1 is under the arc
traced by the tower. Follow the setup procedure described in the
next section to properly calibrate the AutoSampler/AutoInjector
sampling and inject positions.
CAUTIO
If you re-locate the CP-8400/8410 to the rear position, you MUST
recalibrate the injector positions. You will not be automatically
prompted to recalibrate the injector positions. If you fail to
re-calibrate the injector positions, the AutoSampler may miss the
injector, resulting in bent syringe needles
Plunger Retaining Nut
(unscrew to remove
syringe and tighten
when installing)
Syringe Retaining Clip
(lift up to remove
syringe)