EasyManua.ls Logo

Air Techniques MOJAVE LT3 - Daily Vacuum Line Cleaning

Air Techniques MOJAVE LT3
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page
23Air Techniques, Inc. Air Techniques, Inc.
Page
22
MAINTENANCE
Warning: The following steps must be performed while wearing skin and eye protection designed
for handling typical Haz-Mat material. Use care at all times to prevent spillage.
Caution: Use only CleanStream Cleaner to maintain the system vacuum lines. Do not use chlorine
bleach or solutions of sodium hypochlorite to disinfect the MOJAVE system. These materials
may result in damage or destruction of equipment or loss of system performance.
Daily Maintenance - Clean Vacuum Lines
Clean all vacuum lines in the vacuum system with CleanStream Cleaner daily as part of the overall
preventive maintenance program. This helps to maintain the cleanliness of the vacuum lines and
tubing throughout the system. Using the 2.5 liter bottle of CleanStream Cleaner, PN 57850 and the
CleanStream dispenser system, PN 57665.
CleanStream Dispenser System,
Part No. 57665
Part No. 57850
Part No. Description
57850 2.5 Liter Bottle CleanStream Cleaner (125 applications)
57665 CleanStream Dispenser System
1 Bottle with o-rings and caps
1 Saliver Ejector Adapter
1 High-Volume Actuator Adapter
1 Small Suction Hose Adapter
1 Large Suction Hose Adapter
Required - Not Supplied
Procedure. Prepare the cleaning solution and clean the system daily by performing the following
steps.
Note: The CleanStream Dispenser can hold a maximum of 2 liters of solution for cleaning up
to 2 operatories. Mix solution quantity as necessary.
1. Fill the CleanStream dispenser with tap water as applicable;
a. to the line marked 1 L for 1 operatory
b. to the line marked 2 L for 2 operatories
2. Using the 20ml measuring line in the CleanStream Cleaner bottle cap, add the CleanStream
Cleaner concentrate to the dispenser as applicable;
a. for 1 operatory, add 20ml of CleanStream concentrate to the 1 liter of water
b. for 2 operatories, add 40ml of CleanStream concentrate to the 2 liters of water
3. As shown by Figure 11, the interior of the CleanStream dispenser bottle cap is configured with
three holes for the insertion of HVEs and SEs via provided adapters as follows.
a. attach saliva ejector to smallest atomizing adapter
b. attach 1 or 2 high volume ejectors to respective adapter(s)
4. Refer to Figure 11 and place dispenser in the 2-liter vertical position or the 1-liter horizontal
position as necessary.
5. With the vacuum pump on and handpiece valves open, aspirate the CleanStream solution from
dispenser.
6. After each cleansing procedure, disconnect the hand pieces and rinse the dispenser.

Related product manuals