MULTIJET PRO
3.1
-
TEST FLUID
Use only filtered KEROSENE, from a good provenance.
How to use:
After mounting and duly installing the Mutijet PRO, supply the internal reservoir with the Kerosene as follows.
1) Tun on the equipment through the On/Off switch.
DO NOT FORGET THE CORRECT VOLTAGE SET!
2) Place the funnel (included) in one fo the test tubes, activate the Drain system pressing the Drain key .
3) Carrefuly pour the fluid in the test tube. To increase the speed, use all tubes alternating the fillings.
After 1min, the Drain system will be deactivated automatically, if necessary, reactivate it and continue the
procedure until you see in the rear panel (figure below) that the level is close to the maximum.
ONCE
FILLED, DO NOT INCLINE OR MOVE THE MULTIJET PRO SHARPLY. DISPLACE IT IN THE HORIZONTAL
POSITION. FOR TRANSPORTING, REMOVE THE FLUID FROM THE INTERNAL RESERVOIR.
Replace
teh test liquid mounthly or when it loses translucent appearence, doing
this you will be ensuring the long life of your equipment!!
To change this fluid, use a recipient that can contain 1 liter or more. Place it under the outflow nozzle of the test
fluid reservoir (found in the side of the equipment). Remove the cover and wait for the outflow to end.
GE
T USED TO OBSERVE THE REAR PANEL
NEVER MIX ANOTHER TYPE OF FLUID IN THE TEST TUBES OTHER THAN THE ULTRA TEST. THIS
PROCEDURE WILL PREVENT CONTAMINATION OF THE TEST FLUID AND EARLY DAMAGES TO
THE ELECTRIC PUMP.
Fluid level