QC-200-S-005-1706E
AUTOCLAVE
(Steam Sterilizer)
CL- 32S / 32L
40S / 40M / 40L
Instruction Manual
● Make sure to read this manual to use this product safely.
● Place this manual in a convenient location for easy reference.

Questions and Answers

  • S
    Shawna GriffinAug 21, 2025
    How to troubleshoot if the ALP Laboratory Equipment power is not on?
    • K
      Kimberly Berger DDSAug 22, 2025
      If the ALP Laboratory Equipment does not power on, first ensure the power cable is securely connected. If it is already connected, check that it's properly connected. If the issue persists, there may be burning in the connecting part of the power outlet or breaker, in which case you should contact your local distributor.
  • L
    Larry DawsonSep 5, 2025
    Why is there steam leaking from the lid of my ALP CL- 40M?
    • J
      Jaime Reyes PhDSep 6, 2025
      If there's steam leaking from the lid of the ALP Laboratory Equipment, it could be due to deterioration of the lid gasket (cracks or hardening) or misalignment of the lid and chamber. Contact your local distributor to resolve this issue.
  • K
    Kristen ChurchSep 4, 2025
    What does 'Er0' mean on the ALP CL- 40M display?
    • M
      Matthew RichardsonSep 4, 2025
      If 'Er0' is displayed on the ALP Laboratory Equipment, it indicates a memory error of the control unit. Try powering the unit off and then on again. If the problem persists, contact your local distributor.
  • N
    nsanchezSep 2, 2025
    Why is the ALP CL- 40M Laboratory Equipment not reaching the set temperature?
    • T
      Timothy ReyesSep 2, 2025
      If the temperature in the ALP Laboratory Equipment chamber rises, but does not reach the set temperature, check that the electricity supply matches the required voltage of the autoclave and supply proper electricity to the autoclave.
  • H
    hfaulknerSep 1, 2025
    Why is the lid lock lever on my ALP CL- 40M Laboratory Equipment hard to move?
    • K
      Keith StricklandSep 1, 2025
      If the lid lock lever on the ALP Laboratory Equipment is hard to move, it might be because it's being forced when the lid isn't completely closed, or there's insufficient space between the lid and clamps. Try sliding the lid lock lever slowly to the closed position (left end) while pressing the front part of the lid cover, especially if the TEMPERATURE DISPLAY indicates over 60?. If the issue persists, contact your local distributor.
  • K
    Kimberly Dennis DDSAug 30, 2025
    What to do if my ALP CL- 40M Laboratory Equipment beeps and won't start?
    • N
      nicholemurphyAug 30, 2025
      If you hear an error sound and the ALP Laboratory Equipment doesn't start when pressing the START KEY, the completion process of the previous operation may not be reset. Press the STOP KEY to reset the completion process, then try pressing the START KEY again.
  • J
    Jodi MartinAug 29, 2025
    Why isn't sterilization completing in my ALP CL- 40M Laboratory Equipment even at the set temperature?
    • D
      Dennis KiddAug 29, 2025
      If the ALP Laboratory Equipment reaches the set temperature but sterilization isn't completed because the chamber doesn't fill with saturated steam, it could be due to several factors: * Too much capacity of objects in chamber. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. * Shortage of sterilization time. Increase sterilization time appropriately, especially for liquid objects or large quantities. * Residual air inside container prevents penetration of steam. If using disposable plastic bags, ensure they are filled to less than 60% capacity and have an open mouth. * Air evacuation is insufficient due to bent exhaust hose. Make sure the exhaust hose is not bent.
  • L
    Laura JohnsonAug 27, 2025
    Why is the safety valve activating on my ALP Laboratory Equipment?
    • J
      Jeffrey RodriguezAug 27, 2025
      If the safety valve activates on the ALP Laboratory Equipment (steam blowing from the bottom plate), it may be because residual air is not evacuated enough, causing abnormal pressure, or the chamber pressure surpassed set pressure of safety valve due to the malfunction of temperature sensor. Ensure that the quantity of objects inside the chamber is less than 60% of its capacity, with adequate space between them. If the issue persists, contact your local distributor.
  • D
    Douglas DavidAug 25, 2025
    Why does the breaker trip immediately when I turn on my ALP CL- 40M?
    • B
      Bruce SalazarAug 25, 2025
      If the breaker trips immediately after powering on the ALP Laboratory Equipment, it could be due to electrical leakage (possibly from insulation deterioration or water leakage at the heating element) or a short circuit (in the heating element or elsewhere). In any of these cases, contact your local distributor.
  • L
    Leonard MillerAug 23, 2025
    What to do if the lid lock lever on my ALP CL- 40M cannot be released?
    • M
      Michael LopezAug 23, 2025
      If the lid lock lever of the ALP Laboratory Equipment cannot be released, ensure the power switch (breaker) is turned on. If it is on, and the TEMPERATURE DISPLAY indicates above 80° after operation complete or the COMPLETE LAMP lights up after operation, press the STOP KEY to release the lid lock.

Summary

2. For the Safety Use

3. Part Names

4. Installation

5. Operation Method

Basic Operation and Pre-Checks

Step-by-step guide for standard operation, including pre-checks and water supply.

Operation Patterns, Parameters, and Exhaust

Configuring patterns, temperature, time, and exhaust modes for sterilization.

Starting, Completing, and Unloading Operation

Initiating sterilization, post-operation procedures, and safe unloading of items.

6. Safety Devices

Valve, Detector, and Prevention Mechanisms

Functionality of safety valve, over pressure detector, and water lack prevention.

7. Troubleshooting

Error Code Reference

List of error codes and their general meanings for diagnosing issues.

Common Operational Problem Resolution

Guide to diagnosing and resolving common problems like power issues or temperature failures.

8. Maintenance

9. Specifications

ALP CL- 40M Specifications

General IconGeneral
BrandALP
ModelCL- 40M
CategoryLaboratory Equipment
LanguageEnglish