Section 1 • Introduction
22
2.4 AttachCassettetoPump
Insert the cassette onto the top of the pump, as shown. Align and
gently squeeze the tabs on the cassette to attach to the pump.
NOTE: Once the cassette has been properly placed on the
pump,free-ow(unimpededowduetoforcesnot
generated by the pump) cannot occur because the
cassette disc bottom has been snapped into place.
2.5 RemoveCassettefromPump
To remove the cassette, press both cassette release tabs at the same time
and lift the cassette off the pump.
CAUTION: Do not remove the cassette while the green run light is
blinking.Alwaysplacethepumpinpausemoderst.
Failure to do so will cause (1) part of the history to be
lost, (2) possible under-infusion, and (3) the pump to
sound an alarm when the pump attempts to infuse
medication.
NOTE: ThelterbondedtotheambIT
®
cassettetubingisanair-eliminationlter
(“AEF”). The AEF has two membranes. The larger membrane is a hydrophilic
membrane and the smaller membrane is a hydrophobic membrane. The
hydrophobicmembranewillnotallowwatertoowthroughit.However,in
conjunctionwithsomeotherliquids,suchasorganicliquids,watermaybe
permittedtoowthroughthehydrophobicmembrane.
NOTE: Whenthelterisdry,boththehydrophilicmembraneandthehydrophobic
membrane will allow air through. Once the hydrophilic membrane comes in
contactwithwater(is“wetted”)airwillnotowthroughthemembraneuntil
the bubble point is reached. (The bubble point is the pressure required to force
air through the wetted hydrophilic membrane).
NOTE: Due to the nature of the hydrophilic membrane, it is important to prime the
ambIT
®
lteredcassettewithoutgettingwaterinthelter.Thiswillallowairto
owthroughveryeasily.Oncethelteriswet,itmaybemoredifculttoprime,
becausetheairhasamuchsmallerareatoowoutof(airwillonlyowoutof
the smaller hydrophobic membrane).
F
F
ambIT
®
pump
Tubing
on left side
of pump
Larger tab on
right side
Sensor
alignment