SECTION 11 - GENERAL CARE INSTRUCTIONS
WARNING: Pump failure may be caused by the application of cleaning solutions other
than those recommended by the manufacturer. Do not immerse the pump or
sterilize cassette in any cleaning solutions.
The patient should be careful to protect the pump at all times. The pump should not be
dropped.
Transportandstorageconditions:-25ºC(-13ºF)withoutrelativehumiditycontrol;and+70ºC
(+158ºF)atrelativehumidityofupto93%,non-condensing.
The pump will warm from the minimum storage/transportation temperature to room
temperature(about20ºC[68ºF])inapproximately30minutes.Thepumpwillcooltoroom
temperaturefromthemaximumstorage/transportationtemperatureinabout35minutes.
Operatingconditions:+5ºCto+40ºC(+41ºFto+104ºF);relativehumidityrangeof15%to93%,
non-condensing;andanatmosphericpressureof700hPato1060hPa(10.2psito15.4psi).
The pump and components should be stored in a dry, cool place until used.
No sterilization of the pump is required. Disinfect the pump before and after every patient use,
procedure, and/or transfer of patients. (See Section 11.2 for instructions.)
No maintenance of the pump is required, and no calibration is required. Contact Summit
Medical Products, Inc. if a functional test is desired.
NOTE: For storing and transporting the pump, a cap should be placed on the pump to
protect the pressure switch.
11.1 WarrantyInformation
Contactyourlocalsalesrepresentativeforwarrantyandextendedwarrantylengths.
This warranty will not apply to ambIT
®
pumpsthathavebeen,inthejudgmentofSummit
Medical Products, Inc., damaged in whole or in part due to misuse, abuse, negligence,
alteration or improper installation, or that have been dropped or used in a manner inconsistent
with their labeling and packaging.
To obtain warranty service, the pump and cassette must be returned to Summit Medical
Products,Inc.withpostageprepaid.Thereplacementofapumpandcassettewillnotextend
the original term set forth above.