Section 4 • Operating Instructions
46
4.6 SummaryofOperatingControls
ACTION STEPS TO TAKE
AUDIBLE
INDICATOR
VISUAL INDICATOR
Review and verify
program setttings
While the pump is in
pause mode, press and
immediately release the
BOLUS button
One beep “RX” - display cycles through
each programmed parameter
Start infusion Program the pump, then
press and release the
RUN/PAUSE button.
The RUN/PAUSE button
will need to be pushed a
second time to start the
infusion after the required
review has taken place.
One beep Green run light (inside the
BOLUS button) blinks, the “ml”
icon and volume infused is in
the pump display.
NOTE:Ifstartinginfusionafterpausingit,theRUN/PAUSEbuttononlyneedstobepressedonce.
Pause infusion Press and release the
RUN/PAUSE button.
Two beeps
(every four
minutes)
Pause icon ( II )ashesinthe
display;greenrunlight(inside
the BOLUS button) stops
blinking.
Silence alarm Press and release the
RUN/PAUSE button.
Alarm sound
stops
Pause icon ( II )ashesinthe
display;greenrunlight(inside
the BOLUS button) stops
blinking.
Deliver bolus (PCA and
P+P modes only)
Press and release the
BOLUS button.
One beep Green run light (inside the
BOLUS button) double blinks.
NOTE:PatientmayusetheBOLUSbuttononthepumporaremoteBOLUSswitchtorequestabolus.
Deliver dose at the
start of an infusion
after changing settings
or clearing history (PIb
and P+P modes only)
Press and release the
BOLUS button within 60
seconds of starting the
infusion.
One beep Green run light (inside the
BOLUS button) blinks, and
“ml” (volume infused) is in the
pump display.