Section 6 • Patient Lockout
50
SECTION 6 - PATIENT LOCKOUT
Thepumphasbeendesignedwithpatientlockoutfeatures:
WARNING: The steps required for entering the pump’s program mode should not be
revealed to the patient. This prevents unauthorized tampering with the infusion
parameters, and prevents the infusion history from being cleared.
6.1 AccessingLockoutMode
Thepump’slockoutmodeisaccessedbyfollowingthesesteps:
Step#1
Place the pump in pause mode.
Step#2
Enter program mode by pressing and holding both the RUN/PAUSE button and the
BOLUS button for two seconds until one beep is heard and “PRO” appears in the pump
display.
Step#3
Release both buttons and immediately press and release both the RUN/PAUSE button
and the BOLUS button again. Four dash symbols ( - - - - ) will appear in the display.
6.2 ToLockthePump
After accessing lockout mode (see above), enter one lock code number in each space of the
display,startingfromtheright.Thestepsareoutlinedbelow:
Step#1
Therstdigitspaceisashing.
Step#2
Press the RUN/PAUSE button to increase the number or press the BOLUS button to
decrease the number.
Step#3
When the desired number appears in the display, press and hold either the RUN/PAUSE
button or the BOLUS button for two seconds until one beep is heard.
Step#4
Repeat Steps 1 through 3 to program the second through fourth digits.