Section 1 • Introduction
25
Topoweroffthepump:
Step#1
If the pump is in run mode place the pump in pause mode by pushing
the RUN/PAUSE button (see Section 4.2).
Step#2
Rotate the battery cover counter-clockwise until the (O) mark on the
cover lines up with the ( I ) mark on the pump (see illustration at left).
2.8 PumpPowerOnandOff
CAUTION: Always place the pump in pause mode prior to turning the pump off. Failure to
do so may cause the therapy to be delayed and/or history to be lost. The pump
will not sound an alarm if the pump is not placed in pause mode prior to being
turned off.
Topoweronthepump:
Rotate the battery cover clockwise until the ( I ) mark on the cover lines up with the ( I ) mark
on the pump.
NOTE
: After the power-on self-test, the pump will beep twice and go into pause mode.
The clinician can then program the pump or resume current infusion settings.
Power OFF
counter-
clockwise
Power ON