Section 9 • Specifications
55
SECTION 9 - SPECIFICATIONS
Specication Details
Stroke volume bolus 50 microliters
**Bolus/dose delivery rate 210 ml/hr
Volumetric accuracy +/-6%
Interval limit (IL) 0 to 50 ml in 1 ml steps
Basalowrates 0 to 20 ml/hr in 0.1 ml/hr steps
Dose volumes 0 to 50 ml in 0.1 ml steps
Doseintervaltimes(hh:mm) 00:01to24:00inoneminuteincrements
Bolus volumes 0 to 50 ml in 0.1 ml steps
Lockouttimes(hh:mm) 00:01to24:00inoneminuteincrements
Volumes to be infused (ml) 1 to 1000
Pump mechanism Microprocessor controlled rotary-peristaltic
Maximuminfusion/occlusionpressures 25 +/- 12 Psi
Maximumactivationtimeofocclusionalarm
(minimum rate) at minimum occlusion
pressures
4 hours
Maximumactivationtimeofocclusionalarm
(intermediaterate)atmaximumocclusion
pressures
90 seconds
Dimensions/weight
2.16in.X1.4in.X6.875In.(55mmx36mmx175mm)
4.7 ounces (133.2 grams) without batteries
6.4 ounces (181.4 grams) with batteries
Power supply Two AA 1.5V batteries
Battery life (rate dependent) ≥14Days@1ml/hror≥26hrs@20ml/hr
Display, audible alarms, signals and reports
Run indicator light
Bolus infusing
Pause indicator
Occlusion downstream (25 +/- 12 psi)
Cassette not mounted on pump
Low battery
Dead battery
Malfunction
Boluses requested
Infusion complete