Section 3 • Programming Instructions
32
3.2.1.3 MovingBackwardinDisplayWhileProgramming
Theusercanmovebackwardinthedisplay.Thisfunctionsasfollows:
Pushing and holding the BOLUS button held until a beep is heard causes the currently
blinking value to be accepted (i.e., stored in memory) and the value to the left on the
display to start to blink so that it can be changed.
If the leftmost value is blinking and the BOLUS button is pushed and held until a beep
is heard, then the value is accepted (i.e., stored in memory) and the display changes the
precedingparameter.Forexampleiftheleftmostvalueoftheboluslockouttime(hours)
is blinking and the BOLUS button is pushed and held held until a beep is heard, the hours
value is accepted and the display changes to the previous parameter (bolus volume) and
the right most digit blinks (tenths digit).
Ifthebasalrateleftmostdigit(tensdigit:0,1X,2X)isblinkingandtheBOLUSbutton
is held until a beep is heard, then the value is accepted and the pump shows the mode
selected “PCA”.
If the mode (PCA) is blinking and the BOLUS button is held until a beep is heard, the
modeisacceptedandthepumpdisplaystherstparameterforthatmodeandtheleft
most value is blinking.
NOTE: If the mode is blinking, then the BOLUS button functions the same as the RUN/
PAUSE button.