Section 3 • Programming Instructions
41
ACTION
AUDIBLE
INDICATOR
VISUAL INDICATOR
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to enable
review or correction.
NOTE:Themaximumlimitforthelockouttimeisdependentuponthedosevolume,doseinterval
time and/or the bolus volume programmed.
8 Setvolumetobeinfused:
(A) Press and release the RUN/
PAUSE button to increase the
number or press and release the
BOLUS button to decrease the
number.
“ml” is displayed, and bag icon
andrstdigittotheleftash.
(B) When the desired number
appears in the display, press and
hold the RUN/PAUSE button or
the BOLUS button until one beep is
heard to accept the setting.
One beep NOTE:Volumetobeinfusedis
displayed in ml.
(C) Repeat steps 7(A) and 7(B) to
set the digits in the tens and ones
sections.
“ml” is displayed. Bag icon and
respectivedigitashes.
Volume to be infused is
established.
(D) If during programming, you
need to move backwards in the
display to a previous setting, press
and hold the BOLUS button until
one beep is heard.
Each time the BOLUS button
is pressed, the display moves
backward one screen to enable
review or correction.
NOTE:Bagvolumelimitsare1to1,000ml.
Pump returns to pause mode. One beep
followed by
two beeps
Pauseiconashes.
NOTE:Toconrmprogramsettings,access“RX”bypressingandreleasingtheBOLUSbutton.After
scrolling through the programmed settings, the pump will automatically return to pause
mode.
3.2.3.3 MovingBackwardinDisplayWhileProgramming
Theusercanmovebackwardinthedisplay.Thisfunctionsasfollows:
• Pushing and holding the BOLUS button held until a beep is heard causes the currently
blinking value to be accepted (i.e., stored in memory) and the value to the left on the
display to start to blink so that it can be changed.
ml
ml