Section 12 • Electromagnetic Immunity (EMC)
66
Table12-1.ElectromagneticimmunitylevelstestedandpassedbyambIT
®
pump.
Phenomenon Basic EMC Standard Or Test
Method
ImmunityTestLevels
Electrostatic discharge IEC 61000-4-2
±2, 4, 6, 8 kV contact
±2, 4, 8, 15 kV air
RadiatedRFEMelds IEC 61000-4-3
10 V/m
80 MHz-2.7 GHz
Rated power frequency magnetic
elds
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
Table12-2.MaximummeasuredradiatedemissionlevelsfromtheambIT
®
pumpduring
operation.
Frequency(MHz)
dB(µV/m)
Quasi-peak(limit) Actualvalue(pass/fail)
30 to 230 30 23.85 (pass)
230 to 1000 37 29.31 (pass)