2.10.4 Accessing the Calibration Modes
1. On the top side of ThePoint, temporarily remove the shunt
from the “Time Delay Selection Jumper” (see Fig. 2) and
place it on pins 1 & 2 of the 3-pin connector. Pin 1 is closet
to the LEDS. The green LED will go out and the red LED
will begin to ash. The number of ashes indicates which
mode the unit is in (1 through 8).
2. To switch modes, press and hold the ReCal button next to
the 3-pin connector. The unit will cycle through the modes.
First it will ash the current mode setting, then progress
through all of the settings.
For Example:
The red LED will ash once indicating mode 1. Then it will
ash twice-indicating mode 2. Then mode 3, etc.
Release the button when it reaches the desired mode. The
red LED will now ash indicating which mode the unit is
in.
3. Remove the shunt from pins 1 & 2 on the 3-pin connector
and replace the shunt on the “Time Delay Selection Jumper”.
The unit will remain in the selected mode. Write the new
mode # on the inside of the lid label for future reference
4. After setting the mode follow procedure in section 2.8.2
for mode 2 & 6. For modes 7 and 8, follow the appropriate
manual calibration procedure as described in section 2.10.3.
Electronic Unit Adjustments
Do not push
the Recal button
without rst ensuring
that the material
being measured is
below the sensing
element.
Failsafe Jumper
Up = High Level
Down = Low Level
Time Delay Jumper
Up = Forward Acting
Down = Reverse Acting
ReCal Button
3 Pin Connector