Troubleshooting
25
3.3 Testing Relay Circuits
Use the following steps to check out the relay circuits:
1. Relay connections consist of a double-pole double-throw
(DPDT) relay.
2. The relay contacts are brought out to terminal strips for
external switching. See Figure 3-2.
3. Relay operation may generally be heard as an audible click
when background noise is not too high. Connect ohmmeter
to relay contacts to determine if they are switching.
4 6 75 8 9
One 5A DPDT Relay
NOTE:
Relay Shown as Power Off
Figure 3-2
Relay Circuit Operation
3.4 Over Range
If the GREEN LED is ashing, the instrument has detected
the uncovered sensing element capacitance exceeds the limits
of the transmitter. Consult factory instructions.
3.5 Under Range
If the RED LED is ashing, the instrument has detected the
sensing element capacitance is too small. Consult factory for
sensing element capacitor values.