Serial Communications | 5-7
Variables Table
Name ID HEX Description Type Typical Value
Press Units 13 D Process pressure units of measure Integer 0
MaxZeroError 18 12 Maximum zero error allowed Float 0.5
MaxZeroDrift 19 13 Maximum zero drift allowed Float 1.0
AnalogOutSpan 29 1D Analog output span Float 1.0
AnalogOutOffset 30 1E Analog output offset Float 0.0
Baud 40 28 Serial communication speed Integer 9600
SerialMode 41 29 RS-485 two wire or four wire mode Integer 1(four wire)
NodeAddress 42 2A RS-485 analyzer address String 1
AnalyzerName 43 2B Ametek analyzer name String 5830
Gas 44 2C Sample gas name String Nitrogen
Alarm1Hi 46 2E High concentration limit Float 100.0
Alarm1Lo 47 2F Low concentration limit Float 0.0
MaxCalError 55 37 Maximum verifi cation error allowed Float unused
KFactor 56 38 Gas based coeffi cient Float 1.0
CalHour 57 39 Verifi cation schedule parameter Integer 13
CalPeriod 58 3A Verifi cation schedule parameter Integer 15
CalMonthDays 59 3B Verifi cation schedule parameter Integer 1
CalWeekDays 60 3C Verifi cation schedule parameter Integer 1
CalType 61 3D Verifi cation schedule parameter Integer 0
SerialNumber 62 3E Analyzer manufacturing number String 123
CellSerialNumber 63 3F Sensor manufacturing code String 583015
MoistureGeneratorSN 64 40 Moisture generator production code String 020101305012
Version 65 41 Embedded software version String 1.00
HoldOut 69 45 Hold out during zero and verifi cation Integer 1(hold)
VerDurationLimit 71 47 Minimum verifi cation cycles number Integer 15
ZeroOnly 72 48 Start zero only fl ag Integer 1(zero only)
EnableAlarm1 73 49 Enable or disable concentration alarm Integer 1(enable)
DryerDateCode 75 4B Dryer production code String 010080009
DryerPpmHours 76 4C Dryer consumption in 1000 ppm hours Float 20000.00
LCDcontrast 79 4F Display contrast setting Integer 5
LCDbacklight 80 50 Display back light on or off Integer 1(on)
AdjustFreqOffset 81 51 Adjust frequency offset after zero Integer 1(adjust)
ZeroHour 82 52 Zero schedule parameter Integer 13
ZeroPeriod 83 53 Number of cycles to stay in zero Integer 15
ZeroMonthDays 84 54 Zero schedule parameter Integer 1
ZeroWeekDays 85 55 Zero schedule parameter Integer 1
ZeroType 86 56 Zero type(daily,monthly,weekly) Integer 0(daily)
Pressure1 87 57 Sensor pressure value Float 14.7 psi
CellTemp 88 58 Sensor temperature value Float 60.0 degreeC
AnalogOutput 89 59 Current output reading Float 12.0 mA
EETemp 90 5A Electronics enclose temperature Float 55.0 degreeC
Moisture 91 5B Moisture concentration reading Float 20.0 ppm
HeldConc 92 5C Held moisture concentration reading Float 22.0 ppm
Frequency 93 5D Sensor frequency reading Float 1200 Hz
DeltaFrequency 94 5E Wet minus dry reading Float 0.01 Hz
Flow3 97 61 Held and displayed fl ow Float 50.0 sccm
ValveState 99 63 0-sample, 1-ver, 2-reference, 3-zero Integer 0 (sample)
Hour 100 64 Analyzer current time read Integer 13