EasyManua.ls Logo

Ametek 5800 - Pilot Valve Manifold; Removing the Pilot Valve Manifold

Ametek 5800
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-16 | 5830 Moisture Analyzer
Pilot Valve Manifold
The Pilot Valve Manifold assembly is found on the side of the Power Sup-
ply PCB (Figure 7-9).
Removing the pilot valve manifold
1. Using a 7/64” hex wrench, loosen the two captured screws that hold
the pilot valve manifold assembly to the sample manifold plate (Figure
7-6)
2. Disconnect the pilot valve receptacle plugs (Figure 7-9).
3. Use a screwdriver to remove the two (2) screws at the side of the
manifold (Figure 7-12).
The mounting screws for the two manifolds are different size.
2. Detach the tubing from the manifold by holding the tubing retaining
NOTE
Receptacle Plug
Screws
(Use Slotted
Screwdriver)
Figure 7-12.
Model 5830 pilot valve
manifold.

Table of Contents