Modules 29
nVision Operation Manual
Voltage (V) Input
Accuracy
±(0.015% of reading + 0.002VDC)
Accuracy specications include all eects of linearity, hysteresis, repeatability, temperature, and stability within the specied operating temperature range
for one year.
N
ot
e: Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertication period.
Ranges, Resolutions, and Units
Range . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 28VDC
Max Allowable Voltage . . . . . . . . . . . 30VDC
Resolution . . . . . . . . . . . . . . . . . . . . . . .0.001VDC
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Switch Test
Switch Detection
Switch States . . . . . . . . . . . . . . . . . . . . Dry Contact
Closed State Resistance . . . . . . . . . .< 10 Ω
Open State Resistance . . . . . . . . . . . > 10 MΩ

TEMPERATURE RTD100 MODULE INSTRUCTIONS
Your nVision has the ability to measure temperature very accurately if populated with an RTD100 module. With this system you may connect your resistance
temperature detector (platinum RTD) or platinum resistance thermometer (PRT) to the nVision using the provided IP67 rated connector system (P/N: 3953).
Once the sensing element is connected, you may display the temperature reading in your desired unit. The nVision can also measure electrical resistance (Ω)
to help in troubleshooting your resistance based sensing element.
Temperature Coecient of Resistance (TCR)
Your nVision comes pre-loaded with several common RTD sensing element TCR values with the appropriate Callendar-Van Dusen Coecients to convert your
resistance measurement to the appropriate temperature measurement. The available 100Ω platinum RTD TCRs are:
Pt100 (385) Euro
Pt100 (3911) US
Pt100 (3926)
Simply select the desired TCR or coecient values in CrystalControl for use on the nVision. You may also view the Summary page to conrm
you have selected the correct setting for your sensing element.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Crystal Pressure nVision and is the answer not in the manual?

Ametek Crystal Pressure nVision Specifications

General IconGeneral
BrandAmetek
ModelCrystal Pressure nVision
CategoryVoice Recorder
LanguageEnglish

Related product manuals