Zero
To zero the nVision:
•
  Press the (zero) button for at least 1 second while vented to atmosphere until the dashed lines (- - - - -) appear.
To clear the zero value:
•
  Hold the (zero) button for 5 seconds until the display changes from (- - - - -) to the zeroed value, then to (- - -).
You can adjust the Zero Limit at which the (zero) button will display “- -HI- -” in CrystalControl. You can also disable the (zero) button 
entirely, by setting the zero value to a negative number less than -15 psi.
Note:  If you attempt to zero the gauge with more pressure applied than the Zero Limit set in with CrystalControl, the command will be ignored, and “- -HI- -” 
will display.
Note:  You can never zero the BARO sensor.
 !
WARNING:  This gauge can display zero pressure when connected to a source of pressure! Do not rely on the display indication before disconnecting— 
it may not be indicating true pressure. Never disconnect pressure instrumentation without rst relieving system pressure!