SPOT+
2 - 2REST API User Guide
Parameter Read Write Denition and Range Scaling
alarmStatus
4 8
Alarm status byte. Any value > 0 causes the
CMD Out relay connections to the SPOT+ to
be closed if normally open (or vice versa,
depending on instrument conguration).
Individual bits are set to 1 if the alarm
is congured and the alarm threshold is
exceeded:
Bit 0: low ambient temperature
Bit 1: high ambient temperature
Bit 2: low target temperature
Bit 3: high target temperature
Bit 4: low signal% *
None
watchdog
4
Toggles between 0 and 1 every second,
providing a check that the microprocessor is
running.
None
errors
4
Target emissivity for the shorter wavelength
channel calculated by an Application
pyrometer.
Values from 0 to 1.2 reported to 3 decimal
places.
None
* Available on Ratio or Application specic models
** Available on Application specic models.