SPOT+
2 - 1 REST API User Guide
INDEX OF SPOT+
ETHERNET/IP
PARAMETERS
2
2.1 SPOT+ Key Output Values
Please note that all parameters are transmitted as integers. To achieve
sucient resolution, we apply scaling factors as shown in the table. For
example, a temperature of 732.8 °C is transmitted as 7328 and an emissivity of
0.835 is transmitted as 835.
Parameter Read Write Denition and Range Scaling
targetTemp
4 8
Target temperature output, after time
functions and instrument mode calculations
have been applied.
Values 0 to 6500 in °C or °F (depending
on instrument conguration), reported to 1
decimal place.
Over and under-range values are transmitted
as 6553.5 and 6553.4 respectively.
10
emissivity1
4
**
4
*
Target emissivity for the shorter wavelength
channel calculated by an Application
pyrometer.
Values from 0 to 1.2 reported to 3 decimal
places.
1000
emissivity2
4
**
4
*
Target emissivity for the longer wavelength
channel calculated by an Application
pyrometer.
Values from 0 to 1.2 reported to 3 decimal
places.
1000
ambientTemp
4 8
Internal temperature of the instrument.
Values from 0 to 100 °C or 212 °F
(depending on instrument conguration)
reported to 1 decimal place.
10