931003E / 0620 4. FIRMWARE COMMANDS AND RESPONSES
SET_SHAP 2 Sets the shaping time to 1.25 ìs
SET_SHAP 3 Sets the shaping time to 1.5 ìs
SET_SHAP 4 Sets the shaping time to 1.75 ìs
SET_SHAP 5 Sets the shaping time to 2 ìs
SET_TRIG_SAMPLE setting
Selects the triggering source in InSight Mode. See LIST_TRIG_SAMP for legal trigger
sources.
SET_TRUE_PRESET ticks
Sets the true-time preset to the specified number of ticks (20 ms/tick). During data
acquisition when the true-time counter reaches the preset number of ticks, the preset is
complete and the acquisition is stopped. Setting a true-time preset to 0 ticks disables the
preset. The preset should always be set to a multiple of one second. See also
CLEAR_PRESETS and SHOW_TRUE_PRESET.
SET_ULD chans
Sets the upper level discriminator to chans. which must be between 0 and [conversion
gain ! 1]. See also SHOW_ULD.
SET_WINDOW [start, length]
Sets the window of interest to the specified start channel and number of channels. The
channels of spectral data in the window of interest are affected by commands such as
CLEAR, SET_DATA, and WRITE. If neither start nor length is provided, the window is set
to the maximum size allowed by the conversion gain specified. The window of interest is
always set to the maximum size after a SET_DEVICE or SET_SEGMENT command.
SET_ZERO_ADJUSTMENT value
Sets the zero stabilization adjustment to an arbitrary value from !65535 to 65535. The total
range of the adjustment value represents ±256 channels. This adjustment is usually only
made by the gain stabilizer, and reset to 0 with the INITIALIZE_ZERO_STABILIZATION
command. See also SHOW_ZERO_ADJUSTMENT.
SET_ZERO_CHANNEL chan
Sets the center channel for the stabilizer zero peak. If a zero channel is chosen such that the
beginning channel or ending channel would be below channel 0 or above the maximum
channel as determined by the conversion gain, the zero peak width is reduced until the peak
fits the device boundaries.
33